Abstract

A rapid and sensitive flow injection analysis–atomic absorption spectrometric procedure is described for the determination of iron, cobalt, nickel, manganese and zinc based upon on-line preconcentration on a microcolumn packed with C18 material. These metals were complexed with 5,7-dichlorooxine from weakly acidic or neutral solutions in the flow system and adsorbed on the column. The preconcentrated elements were eluted with acidified methanol (pH ≥ 2) and injected directly into the nebulizer for atomization in an air–acetylene flame for measurement. The retention efficiency was better than 98%, resulting in sensitivity enhancement factors of 60, 80, 80, 80 and 60 for a 1 min preconcentration time for iron, cobalt, nickel, manganese and zinc, respectively. The respective detection limits were 4.0, 1.0, 1.0, 0.5 and 0.5 ppb, respectively. The throughput of samples was 30 h–1, with a loading time of 1 min. The method was applied to sea-water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.