Abstract

A self-contained module for liquid—liquid extractions is described. The module contains engraved conduits for mixing of sample and ragent, an engraved segmentor, a detachable extraction coil, a membrane separator, and a rinsing system for the flow cell. The membrane in the separator is supported by a teflon-coated steel grid and can be replaced rapidly. The segmented stream enters at the centre of the circular membrane and travels through an engraved, coiled channel in contact with the membrane before leaving the membrane area at the periphery. The volume of the receptor chamber for the organic phase is 10 μl. For a detector flow-cell volume of 8 μl, an aqueous flow rate of 2.0 ml min −1 and an organic flow rate of 1.2 ml min −1, the “loss factors” caused by analyte dispersion are 3.2 and 1.5 for caffeine sample volumes of 40 μl and 100 μl, respectively, compared with batch extraction. The system is also tested for extraction of anionic surfactants as their ion-pairs with methylene blue.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.