Abstract

A flow injection (FI) method is reported for the determination of thyroxine based on its enhancement of chemiluminescence (CL) from the Ru(bpy) 3 3+–NADH system. The calibration graph was linear over the range 2.0–10 × 10 −8 mol L −1 ( r 2 = 0.9989) with relative standard deviations (R.S.D.) in the range 2.0–4.5% ( n = 4). The limit of detection (3 σ blank) was 1.0 × 10 −9 mol L −1 with sample throughput of 120 h −1. The effect of some organic compounds, anions and cations were studied for l-thyroxine determination. The method was applied to pharmaceutical preparations and the results obtained were in reasonable agreement with the amount labeled. The method was statistically compared with the results obtained by RIA; no significant disagreement at 95% confidence limit was observed. A calibration graph of NADH over the range 1.3 × 10 −8–1.3 × 10 −6 mol L −1 was also established ( r 2 = 0.9992) with R.S.D. in the range1.0–3.5% ( n = 4). The limit of detection (3 σ) was 1.0 × 10 −10 mol L −1 NADH.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.