Abstract
Plumbane produced from the lead analyte in a flow-injection manifold by reaction with sodium borohydride is passed through a porous poly(tetrafluoroethylene) membrane in a gas-diffusion cell. The hydride reacts with bromine in the acceptor stream resulting in ionization which is detected by conductivity measurement. Direct mixing of the carrier with a reagent stream yields a limit of detection of approximately 1 mg/L. An improved detection limit of about 200 microg/L can be achieved by the incorporation of an auxiliary stream containing persulphate as oxidizing agent. The application of the method to the determination of lead in road dust and soil samples is demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.