Abstract

In this paper, a rapid and sensitive flow-injection chemiluminescence (flow-CL) system was developed for the determination of cloxacillin sodium in environmental water samples and pharmaceutical preparations. The method was based on the enhancement effect of cloxacillin sodium on the CL reaction of luminal–H2O2–CuO nanosheets (NSs) in alkaline medium. The CuO nanosheets were synthesized using a green sonochemical method. The physical properties of the synthesized CuO nanosheets were characterized by X-ray diffraction (XRD) and scanning electron microscopy (SEM) analyses. The influences of various experimental factors such as H2O2, NaOH, luminol and CuO nanosheets concentrations were investigated. Under the optimum conditions, the enhanced CL intensity was linearly related to the concentration of cloxacillin sodium in the range of the 0.05–30.00mgL−1 with a correlation coefficient of 0.995. The corresponding detection limit (3σ) was calculated to be 0.026mgL−1. The relative standard deviation (RSD) of the developed method was 2.21% with 11 repeated measurements of 4.00mgL−1 cloxacillin sodium. Also, a total analysis time per sample was 30s which confirmed the rapidity of the proposed method. The analytical applicability of the proposed CL system was assessed by determining cloxacillin sodium in spiked environmental water samples and pharmaceutical preparation. Furthermore, the possible mechanism of CL reaction was discussed.