Abstract

Background The aim of this study was to develop, validate and compare flow injection analysis (FIA) and ultra-high-performance liquid chromatography (LC)/tandem mass spectrometry methods for the determination of imatinib in plasma from patients with chronic myeloid leukemia. Methods The plasma for analysis by both methods was deproteinated by methanol containing d8-imatinib. The separation was achieved on a 1.7 μm C18 column with a linear gradient (4 mM ammonium formiate and acetonitrile, pH 3.2). FIA was performed at flow rate of 0.03 mL/min (0.1% formic acid in methanol). Multiple reaction monitoring mode on the tandem mass spectrometer (API 4000, AB Sciex) in positive ESI were used for detection. Results The total analysis times were 3.2 (LC) and 0.75 min (FIA). Both methods were successfully validated and applied to the plasma patients samples. The limits of quantification were 4.1 and 30.8 ng/mL; imprecisions were less than 5.7% and recovery ranged between 93 and 105%, for the LC and FIA, respectively. The methods revealed an agreement with a mean difference of 1.46 ng/mL (SD 28.95 ng/mL). Conclusions The high-throughput methods that were developed are suitable for the therapeutic drug monitoring of imatinib in plasma. They can be used in routine clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.