Abstract
A new flow injection analysis (FIA) system equipped with an electrochemiluminescent (ECL) detector has been developed and applied for the ECL detection of 2-thiouracil. The FIA-ECL system used a specially designed flow-through ECL thin-layer cell to reduce the dead volume, the IR drop across the cell, and the probability of accumulation of gas bubbles in the cell. It was thus envisioned to improve the detection limit of the FIA-ECL method. After being established, the new FIA-ECL system was used to investigate the ECL response of 2-thiouracil in the presence of the ECL of Ru(bpy)3(2+). It was found that 2-thiouracil could enhance the ECL of Ru(bpy)3(2+) over a wide pH range (pH 4.0-12.0). A highly sensitive method for detection of 2-thiouracil in biological samples was developed by the new FIA-ECL system after optimizing several experimental conditions, such as the applied potential of the working electrode, the pH value of the aqueous solution, the flow rate of carrier solution, and the concentration of Ru(bpy)3(2+).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.