Abstract

Abstract Flow injection method is described for the determination of the antioxidant tert-butylhydroxy anisole (BHA) based on its catalytic oxidation at a nickel hexacyanoferrate (NiHCF) surface modified graphite wax composite electrode fabricated using a new approach. The electrochemical characteristics of the modified electrode were studied using cyclic voltammetry. The voltammetric response of BHA at the modified electrode showed current densities remarkably higher than the bare graphite electrode and occurred at a reduced over voltage of 200 mV. Linear calibration graphs were obtained in the range of 1.2 × 10−6–1.07 × 10−3 M. A detection limit of 6 × 10−7 M was obtained with a correlation coefficient of 0.9983 based on S/N = 3. Reliable results were obtained by applying the proposed flow injection method to determine BHA spiked in dehydrated potato flakes. Results suggest that the developed method can be effectively employed for the determination of BHA from food samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.