Abstract
Abstract A flow injection analysis - inductively coupled plasma/mass spectrometry (FIA/ICP/MS) system is described using a Sciex Elan Model 250 ICP/MS system coupled with a peristaltic pump and injection valve. Sample volumes of 25 μl are injected and samples can be measured at the rate of one per minute. Addition of nitric acid to the carrier generally enhanced and stabilized the recorded signals. Also, the nitric acid tended to lessen interelement effects. Most interelement effects were fairly constant above 0.1 pm added element.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Analytical Letters
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.