Abstract

The Flow Injection technique is shown to provide fast, reliable and sensitive methods for the determination of calcium in various aqueous as well as serum samples; spectrophotometric or potentiometric detection can be used. At sampling rates of 100–110 samples per hour, with 30-μl sample injections, high reproducibility of measurement and low reagent consumption are achieved in both methods. In the spectrophotometric method, the analytical readout is available within 12 s after sample injection at a total reagent consumption of 0.75 ml per analysis. The potentiometric measurement of the calcium activity in serum is placed on a reliable basis by alternating measurements of serum samples and aqueous standards without incurring any non-reproducible changes in potential between aqueous and serum solutions. This permits the simultaneous determination of pH and pCa, the analytical readout being available within XXX s of sample injection. The good agreement between the results obtained with the Flow Injection method and those attained by atomic absorption and EDTA titrations as well as pCa stat-measurements show that the new methods are potentially suitable for routine analysis.