AbstractGold nanotubular electrode ensembles were prepared by using electroless deposition of the metal within the pores of polycarbonate particle track‐etched membranes. Glucose oxidase (Gox), used as a model enzyme, has been immobilized onto preformed self‐assembled monolayers (mercaptoethylamine or mercaptopropionic acid) on electroless gold via cross‐linking with glutaraldehyde or covalent attachment by carbodiimide coupling. Flow‐injection analysis systems in flow‐through or wall‐jet configurations using these Gox‐modified nanoelectrodes are described. The influence of different experimental parameters (i.e., applied potential, flow rate, interferents…︁) on the analytical response of the sensor to glucose has been evaluated. Under optimized conditions, very reproducible results (standard deviations <4%, n=38) were obtained, linear calibration was achieved in the 2×10−4 M to 3×10−2 M concentration range and the detection limit was 2×10−4 M. Moreover, no significant interferences from species like ascorbic and uric acids were observed at a potential of +0.9 V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.