Abstract

Fluid-pumping technology is a mature engineering subject area with a well-documented knowledge base. However, the pump design optimization techniques accepted in industry are geared toward steady-state constant-flow conditions. In contrast, the implantation of a continuous-flow pump to aid the output of the human left ventricle subjects the device to perpetual variation. This study measures pressure-flow performance characteristics for both axial- and centrifugal continuous-flow rotary blood pumps across a wide range of pressure differential values under uniform conditions by means of a novel open-loop flow system. The axial-flow devices show lower hydraulic efficiency. All pumps yield best efficiency point at a head to flow coefficient ratio of approximately 1.7. The open-loop flow system accounts for the dynamic changes associated with human heart physiology and allows for more precise characterization of existing heart pumps and those in development.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.