Abstract
Graft wall pulsation amplitude sensing can provide a measure of functional status, e.g. in hemodialysis access grafts. Current implantable graft monitoring sensors require graft modification and direct bloodstream contact. We propose a new class of piezoresistive flexible pulsation sensors which can be wrapped around the graft to measure wall movement. These sensors must be highly flexible to prevent graft constriction; typical strain sensors are too rigid and the strain sensing range is too limited for this application. We describe a novel additive manufacturing (AM) method for printing polydimethylsiloxane (PDMS) with an internal porous structure, such that material compliance may be tuned anisotropically for a given sensor geometry. Prototype flexible pulsation sensors (FPS) consisting of structured PDMS with an embedded conductive PDMS sensor layer were fabricated and tested. Initial tests demonstrated reliable sensor response to 1-Hz cyclic elongation of 20%, and a sensor gauge factor of 1.0.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of IEEE Sensors. IEEE International Conference on Sensors
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.