Abstract

This paper presents the development of a non-contact, nonintrusive wrist pulse sensor based on the near-field variation of an array resonator. A compact resonator and its array were designed and fabricated on flexible substrate. The reflection coefficient of the resonator can vary as a function of the distance between the resonator and the walls of the major arteries, and the corresponding variation is utilized to obtain heart rate information at the wrist. To detect very weak pulse signals from the main arteries, a sensitivity enhancement technique was devised using a radio frequency (RF) array resonator. The sensor system was implemented with an RF switch to combine or select appropriate signals from the resonator element and was tested using the 2.4 GHz ISM band. The results demonstrated the sensor system's excellent performance in both sequential and simultaneous detection schemes. The measurement results showed that a heartbeat pulse can be detected from both radial and ulnar arteries via the array resonators. Considering the high sensitivity and characteristics, the proposed detection system can be utilized as a wearable, long-term health monitoring device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call