The use of flexible titanium intramedullary nails for management of pediatric long bone fractures and some adult fractures has become common. Nail removal after union can be challenging and often requires a larger exposure than nail placement to allow the insertion of grasping devices, such as pliers. A percutaneous technique for the removal of flexible intramedullary nails using extraction bolts from a broken screw removal set is presented. The technique is especially useful when the end of the nail is not prominent and where a hollow reamer from the same set can expose the nail end for the extraction bolt to be used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.