Abstract
Hypoxia refers to insufficient oxygen amounts at the tissue level unable to maintain adequate homeostasis. Severe hypoxia may occur in the absence of subjective breathlessness due to respiratory failure. Precise monitoring of low blood oxygen saturation is crucially desired, catering to the clinical requirements. However, current pulse oximeters cannot function well in monitoring peripheral oxygen saturation limited by the weak peripheral blood circulation at a low oxygen level. In this work, we propose a flexible hybrid electronic (FHE) with a compact structure and high sensitivity for conveniently monitoring hypoxia. This FHE is composed of 10-µm thickness semiconductors with different materials, functionalities, and sizes. Its performance is demonstrated by monitoring arterial blood oxygen saturation (SaO2) at the body's different arteries. The absolute error is less than 2% within a SaO2 ranging from 99% to 63%. The efficient techniques presented in this work may bring light to the next-generation flexible hybrid electronics and provide potential widespread use in research and clinical applications, especially for emergency treatment.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE transactions on biomedical circuits and systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.