Abstract
AbstractA novel flexible radio frequency (RF) sensor is designed to facilitate the accurate testing of various samples used in the biomedical industry at the industrial, scientific and medical (ISM) frequency band. The proposed RF biosensor comprises a liquid channel-loaded interdigitated capacitor, which is integrated on a coplanar waveguide structure. The prototype of the sensor is fabricated on a 0.13 mm thin biodegradable polyethylene terephthalate polyester film to perform the testing of various bio-graded samples by recording the corresponding resonant frequency. It is observed that there is a noticeable change between the measured resonant frequencies of these samples, which primarily occurs due to the difference in their dielectric properties. The designed sensor was used to monitor and investigate the quality of glycerol, which is the most commonly used raw ingredient in the biomedical and food industry. The determination of glucose concentration in base fluids is considered to ease the challenges faced by doctors and biochemists regarding the monitoring of glucose concentration. It is found that the proposed sensor can quantify the glycerol purity up to the minimum specified adulteration level of 2 and 1% corresponding to toxic contaminants diethylene glycol and ethylene glycol, respectively, and the glucose concentration of 0.5 mg/ml.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Microwave and Wireless Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.