Abstract

Integrated dose sensing in Flat Detectors allows a during pulse control of the X-ray illumination without the need for external dose sensing devices. Standard designs of Flat Detectors do not allow during pulse dose sensing since the information is collected from the pixels only in the read-out phase after the X-ray illumination. This paper introduces a special detector plate design for obtaining dose sensing information directly from the X-ray detector while the X-ray pulse is being applied. This dose sensing information is read at a lower spatial resolution than the actual X-ray image but with a sub-millisecond temporal resolution. The dose sensing operates without any additional radiation burden on the patient and without attenuation of the image information. Experimental results from a small area (4x4 cm2) detector are presented, including an analysis of noise, linearity and cross-talk.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.