Abstract
A flash x-ray tube has been fabricated and has been driven by a Blumlein pulse generator to peak voltages of 60 to 80 kV and peak anode currents of 8 to 10 kA. The pulse width, dependent upon the anode-cathode spacing, can be made ∼80-ns wide, and a small spot or line focus is possible. These features make the device useful for flash x-ray diffraction studies. Current in the diode region exceeds the calculated space-charge limited values, probably because of neutralization of charge by plasma ejected from the cathode during the initial phase of the discharge.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.