Abstract

In the present work, a new ligandless-ultrasound-assisted emulsification microextraction (LL-USAEME) method was developed for preconcentration trace amounts of nickel as a prior step to its determination by flame atomic absorption spectrometry. Some factors influencing the extraction efficiency of nickel and its subsequent determination were studied and optimized, such as type and volume of the extraction solvent, pH, extraction time, extraction temperature and ionic strength. Under the optimal conditions, the calibration curve was linear over the range of 1.0 ng mL(-1) - 1.0 µg mL(-1) for nickel with R(2) = 0.9997. The detection limit was 0.34 ng mL(-1) in the original solution (3S(b)/m) and the relative standard deviation for 8 replicate determination of 0.3 µg mL(-1) nickel was ±1.6%. The proposed method was successfully applied in the analysis of four real environmental water samples and good spiked recoveries over the range of 97 - 103% were obtained.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call