Abstract
AbstractIn the present article, a procedure for the simultaneous separation and preconcentration of trace amounts of cadmium and zinc is proposed. It is based on the adsorption of cadmium and zinc ions onto a column of Amberlite XAD‐4 resin loaded with aluminon reagent. Cadmium and zinc ions are quantitatively retained on the column in the pH range from 6.5–7.5, at a flow rate of 2 mL min–1. The cadmium and zinc ions are eluted with 5.0 mL of 5 mol L–1 HNO3 solution. Cadmium and zinc are measured by flame atomic absorption spectrometry (FAAS). In the present case, 0.1 μg of cadmium and 0.5 μg of zinc can be concentrated in the column from 1000 mL of aqueous sample, where their concentrations are as low as 0.1 and 0.5 ng mL–1, respectively. The relative standard deviations, for seven replicated determinations of 1.0 μg mL–1 of cadmium and zinc, are 1.2 and 1.1%, respectively. The detection limits for cadmium and zinc in the original solution are 0.02 and 0.11 ng mL–1, respectively. The interference of a large number of anions and cations has been studied and the optimized conditions are utilized for the determination of trace amounts of cadmium and zinc in different environmental and standard samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.