Abstract

A simple and sensitive flame atomic absorption spectrometry technique was developed for the determination of trace amounts of cobalt (Co) in various samples after adsorption on modified analcime using a recently synthesized tetradentate Schiff base ligand, bis-(2-hydroxy-1-naphthaldimine)-N-diethylene-triamine, by a column method in the pH range of 5 to 7, with a flow rate of 1 mL/min. The influence of flow rates of the stripping and sample solution, the amount of ligand, types of stripping, and the least amount of eluent required for elution of the Co ion from the column were investigated. The effects of interfering ions on the recovery of Co were studied. The method was successfully applied for determination of Co in water and plant samples. The maximum capacity of the modified sorbent was found to be 0.58 mg/g of Co(II); the preconcentration factor was 250. Linearity was maintained at 0.2-40 microg/L in the original solution. The LOD was 1.44 x 10(-2) microg/L in the original solution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.