Abstract

The objective of this study was to evaluate the fixation strength of a biodegradable interference screw (Arthrex, Naples, FL) compared with press fit fixation and a titanium interference screw in anterior cruciate ligament (ACL) reconstruction using a bone-patellar tendon-bone (BPTB) graft. Porcine lower limbs were used. Ultimate failure loads of the biodegradable screw (805.2 N; range 680 to 995 N) did not differ significantly from ultimate failure loads of titanium interference screws (768.6 N; range 544 to 1094 N). Press fit fixation provided significant (P < .01) lower fixation strength (462.5 N; range 80 to 825 N). These results support the conclusion that biodegradable interference screws are a reasonable alternative in terms of primary fixation strength.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.