The aim of this study was to compare the postoperative fixation properties of a biodegradable osteosynthesis "free-form" plate achieved with countersunk screws with those provided by screws with cut-off screw heads. Acrylic pipes were fixed together to simulate fracture fixation for tensile testing. Additional plates were fixed to a polyurethane block with a single screw for plate-screw pullout testing. Specimens were incubated in phosphate buffer solution at 37 degrees C, and testing was conducted at various time points during hydrolytic degradation of 26 weeks. In both tests the specimens were loaded at a speed of 5 mm/min until failure. The yield load, maximum load, and stiffness were recorded, and failure mode was visually determined. Both countersunk screws and screws with cut-off screw heads provided similar plate fixation properties over degradation time. According to these results, fixation of the biodegradable osteosynthesis free-form plate with screws with cut-off screw heads seems to be feasible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.