Abstract

We used an experimental model producing a tissue response with a sclerotic endosteal neo-cortical rim associated with implant loosening in humans: a 6 mm PMMA cylinder pistoned 500 m concentrically in a 7.5 mm hole, with polyethylene particles. At a second operation at eight weeks, the standard revision procedure removed the fibrous membrane in one knee, and the crack revision procedure was used to crack the sclerotic endosteal rim in the contralateral knee. Once stability was achieved following the revision procedures, loaded Ti plasma sprayed implants were inserted into the revision cavities of 8 dogs for an additional 4 weeks. Revision implant fixation (ultimate shear strength and energy absorption) was significantly enhanced by cracking the sclerotic endosteal rim. In conclusion, we demonstrated a simple technique of cracking the sclerotic endosteal rim as an additional method for improving revision fixation. (Hip International 2002; 2: 77-9).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.