Abstract

In the 18 months up to July 1983, 120 Björk-Shiley convexo-concave prostheses with 70 degrees opening angle were implanted in 47 women and 61 men aged 19-78 (mean 58.6) years. The prosthetic valves were aortic in 65 cases, mitral in 23 and both aortic and mitral in 20 cases. Emergency operation was required in ten cases, and concomitant surgery was performed in 43 (39.8%). The early mortality was 5.5%. A follow-up study, comprising 498 patient years, revealed 73.1% 5-year survival. There were three mechanical failures of prosthesis, in one of which re-replacement was successful. Elective prosthesis replacement was performed in four other cases judged to be at high risk of strut fracture.