Abstract

Before introducing the transobturator tape into our practice we undertook 52 transobturator tape procedures. The transobturator tape procedures were undertaken with an "outside-in" approach, using nonwoven polypropylene mesh with average pore size of 50 mum. Five cases of vaginal erosions have been identified, 1 complicated by a groin abscess. All cases required further procedures to trim (n = 3), resect (n = 1) or remove (n = 1) the tape. One woman had a tension-free vaginal tape procedure. To date, 3 women remain incontinent. Possible reasons for the complications include 1) surgical inexperience (unlikely, given that we have undertaken more than 2000 tension-free vaginal tape procedures without similar complication rates); 2) inherent susceptibility of the "hammock" position of the transobturator tape; or 3) the nonwoven polypropylene tape with mesh size of 50 microm itself may predispose to erosion or abscess.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.