Abstract

This manuscript describes a neutron counting system based on cylindrical fission counters that can monitor neutron activity for high-energy neutron flux above 10 MeV under electrically noisy environments with intense gamma rays. Miniature fission counters with depleted uranium as sensitive material and modular electronics were built for digital signal processing and high-countrate operation. The counters are 9.5 mm in diameter and 71.1 mm in active length. The author presents the results of Monte Carlo simulations of the fission-counter response for selected neutron sources and energies based on ENDF7.1, JENDL-5, and TENDL-2021 nuclear data libraries from 1 meV to 200 MeV. For a white neutron beam (E‾ = 16.36 MeV) that irradiates the front face of a counter, the intrinsic efficiency is evaluated to be (2.24 ± 0.02) × 10−5 counts/n, while the efficiency of the counter in the array appears to increase by at most 6.7%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call