Abstract

The operational problems of the gamma ray spectrometer shielded passively with 12 cm of lead and actively by five 0.5 m × 0.5 m × 0.05 m plastic veto shields are described. The active shielding effect from environmental gamma ray, cosmic muons, and neutrons was investigated. Wide range of scintillator pulses, corresponding to the energy range of 150 keV–75 MeV, was used for anticoincidence gating. With the optimal set up, the integral background of 0.31 c/s was achieved for the energy region of 50–3000 keV. The detector mass-related background was 0.345 c/kg s. The 511 keV annihilation line was reduced by factor of 7 by the anticoincidence gate. It is shown that the plastic shields increase the neutron capture gamma line intensities due to neutron termalization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.