Abstract

Acute increase of intracranial pressure (ICP) usually has to be treated with an external ventricular drainage (EVD). Current standard mechanical EVD carry a lot of disadvantages, which hypothetically could be bet- ter managed by a newly developed electromechanical EVD. In this report our first preliminary results of such an elec- tromechanical EVD applied in a porcine animal model are presented. The drainage was demonstrated to be both suc- cessful in monitoring and controlling elevated ICP, and able to detect slit ventricles due to overdrainage, if the indented target ICP was set too low.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call