Abstract

We report a case of a late type III endoleak from a hole in the fabric of the main body of a Zenith bifurcated endograft 7 years after implantation. Abdominal pain and a rapidly expanding aneurysm were eventually followed by rupture. The defect was detected at open surgery, whereas no evidence of endoleak was found at preoperative computed tomography (CT) or angiogram. The defect was repaired by a relining procedure with an Excluder stent graft. The patient, however, died 3 weeks after admission.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.