Abstract

Proton minibeam radiation therapy (pMBRT) is a novel dose delivery method based on spatial dose fractionation. pMBRT has been shown to be promising in terms of reduced side effects and superior tumour control in high-grade glioma-bearing rats compared to standard irradiation. These findings, together with the recent optimized implementation of pMBRT in a clinical pencil beam scanning system, have triggered reflection on the possible application to patient treatments. In this context, the present study was designed to conduct a first theoretical investigation of the clinical potential of this technique. For this purpose, a dedicated dose engine was developed and used to evaluate two clinically relevant patient treatment plans (high-grade glioma and meningioma). Treatment plans were compared with standard proton therapy plans assessed by means of a commercial treatment planning system (ECLIPSE-Varian Medical systems) and Monte Carlo simulations. A multislit brass collimator consisting of 0.4 mm wide slits separated by a centre-to-centre distance of 4 or 6 mm was placed between the nozzle and the patient to shape the planar minibeams. For each plan, spread-out Bragg peaks and homogeneous dose distributions (±7% dose variations) can be obtained in target volumes. The Peak-to-Valley Dose Ratios (PVDR) were evaluated between 9.2 and 12.8 at a depth of 20 mm for meningioma and glioma, respectively. Dose volume histograms (DVHs) for target volumes and organs at risk were quantitatively compared, resulting in a slightly better target homogeneity with standard PT than with pMBRT plans, but similar DVHs for deep-seated organs-at-risk and lower average dose for shallow organs. The proposed delivery method evaluated in this work opens the way to an effective treatment for radioresistant tumours and will support the design of future clinical research.

Highlights

  • Proton minibeam radiation therapy is a novel dose delivery method based on spatial dose fractionation. pMBRT has been shown to be promising in terms of reduced side effects and superior tumour control in high-grade glioma-bearing rats compared to standard irradiation

  • Good coverage of the planning target volume (PTV) by the 90% isodose was observed for treatment plans 1 to 3, while heterogeneities were observed for treatment plan 4

  • For a ctc of 4 mm, the Peak-to-Valley Dose Ratios (PVDR) evaluated in the PTV region was 1.04 ± 0.12 for the glioma and 1.05 ± 0.12 for the meningioma

Read more

Summary

Introduction

Proton minibeam radiation therapy (pMBRT) is a novel dose delivery method based on spatial dose fractionation. pMBRT has been shown to be promising in terms of reduced side effects and superior tumour control in high-grade glioma-bearing rats compared to standard irradiation. PMBRT has been shown to be promising in terms of reduced side effects and superior tumour control in high-grade glioma-bearing rats compared to standard irradiation These findings, together with the recent optimized implementation of pMBRT in a clinical pencil beam scanning system, have triggered reflection on the possible application to patient treatments. Equivalent or superior tumour control to that obtained by standard proton therapy has been observed in glioma-bearing rats[5,6] These results, together with the recent implementation of pMBRT in the Orsay proton therapy centre (ICPO) pencil beam scanning system[7], have triggered reflection on the possible evaluation of this technique in phase I/II clinical trials. To the best of our knowledge, this study represents the first complete dosimetric evaluation in human patients for such small proton field sizes using a clinical setup

Methods
Results
Conclusion