Abstract

MedAustron is a synchrotron-based Particle Therapy Accelerator located in lower Austria which delivers clinical proton and carbon beams in the range of 62-252.7 MeV/u and 120-400 MeV/u respectively in two clinical treatment rooms. A proton Gantry was recently commissioned in a third clinical treatment room and a fourth experimental beamline is dedicated to non-clinical research activities. Within the latter, the injector commissioning has recently started for He2+ beam generation. The long-term goal is to use helium for cancer treatment due to its favourable physical and biological properties. The helium beam properties from the source up to injection further downstream into the accelerator, i.e. from the Low Energy Beam Transfer line to the Linear Accelerator are discussed in terms of beam emittance, intensity and transmission efficiency. Furthermore, a comparison with simulated data is presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.