Abstract

PurposeTo report the technology and preliminary result of gutter-free design stent-grafted in in situ fenestration thoracic endovascular aortic repair (TEVAR).DescriptionThe gutter-free stent-graft has a nickel-titanium self-expanding skeleton, double polytetrafluoroethylene coating, and an outer-skirt fabric structure (named C-skirt endograft). The outer skirt fabric prevents endoleak from the gutter around the stent graft fenestration. Further, the skirt structure right under the fenestration in the aortic stent graft can function as a fixation of the side-branch artery endograft. These designs have the following advantages, such as: 1) prevention of endoleak; and 2) fixation tightly between the branch and aorta endograft pieces.EvaluationA patient who was diagnosed with an aortic arch aneurysm, combined with localized dissection, has successfully implanted the aortic stent graft and C-skirt endograft for the left subclavian artery. The 6-month follow-up result of the C-skirt in situ fenestration TEVAR is satisfactory without obvious endoleak.ConclusionsThe new gutter-free C-skirt stent graft is being safely and effectively used for aortic arch TEVAR. Long-term evaluation of safety, effectivity, and durability needs to be proven by future multi-center studies.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call