Abstract

This feasibility study tested the capability of high frequency stimulation (HFS) to block muscle contractions elicited by electrical stimulation of the same nerve proximally. During a tendon lengthening surgery in the forearm, the anterior interosseous nerve (AIN) was exposed. A specialized nerve cuff electrode was placed around the nerve, and a stimulating probe held on the nerve 1 cm proximal to the cuff electrode delivered pulses of current causing the pronator quadratus muscle to contract. Through the cuff electrode, 20 kHz HFS was delivered to the nerve for 10 seconds during proximal stimulation. HFS amplitudes between 5 and 10 mA peak-to-peak were tested to determine which produced complete and partial block of the electrically induced contractions. The minimum HFS amplitude that produced complete block was 8 mA, with lower amplitudes producing partial block. In all trials, muscle contractions resumed immediately after HFS was turned off. This demonstration of high frequency electrical nerve block is a milestone in the road to clinical implementation of HFS mediated motor block for spasticity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.