Abstract

RED-100 is a two-phase detector for study of coherent elastic scattering of reactor electron antineutrinos off xenon atomic nuclei. The detector contains a total of 200 kg of liquid xenon in a titanium cryostat with 160 kg of xenon in active volume inside a Teflon-made light collection cage associated with electrode system. The active volume is viewed by two arrays of nineteen 3′′-diameter Hamamatsu R11410-20 PMTs assembled in two planes on top and bottom. The electrode system is equipped with an electron shutter (a patented device) to reduce a “spontaneous” single-electron noise. The detector was tested in a ground-level laboratory. The obtained results demonstrate that detection of coherent elastic scattering of reactor antineutrinos off xenon nuclei at Kalinin nuclear power plant with the RED-100 detector is feasible with a threshold of 4 ionization electrons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.