Abstract

Stent-assisted coiling of wide-necked intracranial aneurysms has become an established treatment and has significantly benefited from the introduction of compliant, self-expanding devices, such as the Enterprise VRD (EP-VRD). We report our experiences with the successor model, the Enterprise2 (EP2) stent in stent-assisted coiling as well as in the treatment of atherosclerotic stenosis. In 11 consecutive patients 12EP2 were used to treat 9 intracranial aneurysms and 2 stenoses. All stents could easily be delivered including partial or complete recapturing when necessary. In two cases with sharp angled curves in the carotid siphon there was kinking and flattening of the stent resulting in incomplete wall apposition of the stent. Moreover, when vascular anatomy showed curves with angles >50° it was regularly observed that the proximal stent markers were asymmetrically arranged along the vessel circumference without influence on the stent apposition. Both findings could be reproduced in asilicone flow model. The EP2 performed well in our small patient cohort; however, above acritical acute angle there may be incomplete wall apposition of the stent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.