Abstract

Malfunction of a prosthetic mitral disc valve was clinically suspected on the basis of absent valve closing sound and an early opening click in a patient with first-degree atrioventricular (A-V) block. Premature closure and opening of the prosthetic valve was demonstrated by echocardiography. All abnormal auscultatory, phonocardiographic, and echocardiographic findings disappeared promptly upon spontaneous restoration of normal A-V conduction. The possible hemodynamic mechanism is discussed, and attention is drawn to this conduction abnormality as a cause of erroneous diagnosis of prosthetic mitral valve malfunction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.