Abstract
The first comparison of spectral responsivity scales in the soft x-ray region from 360 eV to 5000 eV has been carried out between the Physikalisch-Technische Bundesanstalt (PTB) and the National Institute of Advanced Industrial Science and Technology (AIST). Two semiconductor photodiodes used as transfer detectors were calibrated using monochromatized synchrotron radiation and cryogenic electrical substitution radiometers as the primary detector standards at these institutes. The primary detector standard at AIST was confirmed to be in agreement with that at PTB at the level of the relative expanded uncertainty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.