Abstract

The telemetric shunt-integrated sensors available up to now do not deliver negative CSF pressure values. Therefore, a new sensor was designed which is incorporated into the shunt line proximally to the valve. The monitor calculates the pressure values in relation to the atmospheric pressure obtained by a built-in barometer. The sensor covers a pressure range of —100 cmH20 to 200 cmH20 with a resolution of ±7 cmH20. First follow-up experiences are reported with seven patients over 14-17 months. Depending on the patient's body position telemetrically measured CSF-pressures varied between —20 cmH20 in erect and 15 cmH20 in supine position. In a patient with intermittent shunt dysfunction and stiff ventricular walls telemetrically measured CSF values corresponded earlier and were more sensitive with the clinical symptomatology than with the CT scan. So far, recognizable advantages of the telesensor are an improved assessment of shunt dysfunction and marginal CSF pressure increases, with practicability and simplicity of measurements in outpatients and a reduction of expensive CT or MRI controls. [Neurol Res 1999; 21: 117-120]

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.