Abstract

Silicon Carbide (SiC) has been recently proposed and patented as a material for the realization of radiation dosimetry detectors. SiC properties, including its high sensitivity, radiation damage resistance and its extremely small sensitivity region, make it an amazing device susceptible to radiation dosimetric applications of direct ionizing radiations. In this work, a dosimetric characterization of a 10 μm thick SiC detector, irradiated with 62 MeV monochromatic and clinical SOBP proton beams, is reported. Short term reproducibility and linearity with the absorbed dose demonstrate that SiC can be profitably used as radiation dosimetry in proton therapy applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.