Abstract

Recent studies have evaluated the capability of plastic scintillation (PS) as an alternative to liquid scintillation (LS) in radionuclide activity determination without mixed waste production. In order to complete the comparison, we now assess the extent to which PS can be used to quantify mixtures of radionuclides and the influence of the diameter of the plastic scintillation beads in detection efficiency. The results show that the detection efficiency decreases and the spectrum shrink to lower energies when the size of the plastic scintillation beads increases, and that the lower the energy of the beta particle, the greater the variation takes place. Similar behaviour has been observed for beta–gamma and alpha emitters. Two scenarios for the quantification of mixtures are considered, one including two radionuclides ( 14C and 60Co) whose spectra do not overlap significantly, and the other including two radionuclides ( 137Cs and 90Sr/ 90Y), where the spectra of one the isotopes is totally overlapped by the other The calculation has been performed by using the conventional window selection procedure and a new approach in which the selected windows correspond to those with lower quantification errors. Relative errors obtained using the proposed approach (less than 10%) are lower than those of the conventional procedure, even when a radionuclide is completely overlapped, except for those samples with extreme activity ratios that were not included in the window optimization process.