For the first time, a digital Mirror Langmuir Probe (MLP) has successfully sampled plasma temperature, ion saturation current, and floating potential together on a single probe tip in real time in a radio-frequency driven helicon linear plasma device. This is accomplished by feedback control of the bias sweep to ensure a good fit to I-V characteristics with a high frequency, high power digital amplifier, and field-programmable gate array controller. Measurements taken by the MLP were validated by a low speed I-V characteristic manually collected during static plasma conditions. Plasma fluctuations, induced by varying the axial magnetic field (f̃ = 10Hz), were also successfully monitored with the MLP. Further refinement of the digital MLP pushes it toward a turn-key system that minimizes the time to deployment and lessens the learning curve, positioning the digital MLP as a capable diagnostic for the study of low radio-frequency plasma physics. These demonstrations bolster confidence in fielding such digital MLP diagnostics in magnetic confinement experiments with high spatial and adequate temporal resolution, such as edge plasma, scrape-off layer, and divertor probes.