Abstract

The derivative spectrophotometric method was established for simultaneous determination of dexamethasone and cytarabine (cytosine arabinoside). Measurements were made in the zero-crossing wavelengths at 268.0 nm (first derivative) and 264.0 nm (third derivative) for determining dexamethasone and cytarabine, respectively. The calibration graphs were linear in the concentration ranges 0.10 to 10 μg/mL of dexamethasone and 0.25 to 50.0 μg/mL of cytarabine. The limits of detection 0.08 and 0.10 μg/mL and relative standard deviations 3.0 and 1.0% were obtained for dexamethasone and cytarabine, respectively. The possible interfering effect of other substances was also studied to investigate selectivity of the developed method. The proposed method was applied satisfactorily for the simultaneous determination of both drugs in the pharmaceutical formulation and biological fluid samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.