Abstract

NASA's Cosmic Background Explorer (COBE) is designed to investigate the Cosmic Background Radiation (CBR), that permeates the universe as a consequence of the Big Bang. This 3 degree Kelvin radiation is a fossil that contains much information about the early universe. The Far Infrared Absolute Spectrophotometer (FIRAS), will investigate the spectral isotropy of this ancient remnant and look for clues as to the subsequent evolution of the universe. The instrument is a cryogenically cooled, modified Michelson interferometer which operates in the 1 cm to 100 micron wavelength range. FIRAS is designed to provide absolute spectral information, therefore, all possible perturbations to the instrument response must be investigated to minimize distortions of the data. This paper discusses the methodology and resultant variations in the instrument performance noted during room temperature, and liquid nitrogen, (LN2) temperature vibration qualification. Reference alignment shifts in critical components such as the instrument wire-grid beamsplitter are correlated to changes in the instrument spectral response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.