Abstract

Investigation of morphology and mechanical properties of biological specimens using atomic force microscopy (AFM) often requires its operation in liquid environment. Due to the hydrodynamic force, the vibration of AFM cantilevers in liquid shows dramatically different dynamic characteristics from that in air. A good understanding of the dynamics of AFM cantilevers vibrating in liquid is needed for the interpretation of scanning images, selection of AFM operating conditions, and evaluation of sample's mechanical properties. In this study, a finite element (FE) model is used for frequency and transient response analysis of AFM cantilevers in tapping mode (TM) operated in air or liquid. Hydrodynamic force exerted by the fluid on AFM cantilevers is approximated by additional mass and hydrodynamic damping. The additional mass and hydrodynamic damping matrices corresponding to beam elements are derived. With this model, numerical simulations are performed for an AFM cantilever to obtain the frequency and transient responses of the cantilever in air and liquid. The comparison between our simulated results and the experimentally obtained ones shows good agreement. Based on the simulations, different characteristics of cantilever dynamics in air and liquid are discussed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call