Abstract

For the accurate measurement of small masses and forces in a traceable manner to the new definition of the kilogram, a voltage balance apparatus that uses electrostatic force between coaxial electrodes has been developed at the National Metrology Institute of Japan (NMIJ). In the voltage balance measurements, the coaxial electrodes of the apparatus are mechanically aligned by measuring the capacitance to achieve accurate small mass measurements. In this work, we investigated the effect of surface roughness on mechanical alignment of coaxial electrodes for voltage balance measurements using the finite element method (FEM). The results indicate that misalignments caused by surface roughness alter the capacitance gradient but do not significantly increase unwanted lateral forces in the mass measurement compared to that in the ideal condition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.