Abstract

Purpose: To estimate the theoretical corneal refractive error induced by mechanical weakening effects from posterior lamellar keratoplasty (PLKP) in the human cornea. Methods: The refractive effects of PLKP are simulated by finite-element modeling (FEM) as a mathematical function of the thickness of the excised posterior lamellar corneal button, with a nonlinear formulation of stress-strain relation for the corneal material. A theoretical nomogram was developed to correlate the refractive changes to button thickness. Results: The predicted refractive change after PLKP is less than 1 dpt for a 170-µm thickness posterior corneal button over a broad range of Young’s modulus.Thicker buttons result in greater surgically induced refractive errors. Conclusions: According to FEM analysis, the excision of a posterior lamellar button of less than 170 µm thickness produces a minimal predicted refractive change (<1 dpt) in the cornea after PLKP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.