Abstract

This workstudies the impingement on the rim of acetabular liner due to wear on the surface of acetabular liner using finite element simulation. A three dimensional contact model between a femoral head and an acetabular liner was developed. There are three steps in this simulation, i.e. creating the virtualwear on the surface of acetabular liner, applying the load at the femoral head, and rotating the femoral head from neutral position till the impingement occurrence. The virtualwear is created based on the data of wear depth which was obtained from literature. Results showed that the wear on the acetabular liner surface wouldaffected the impingement occurrence, in which the impingement angle becomes narrow. In addition, the failure possibility of the acetabular liner rimwould become higher.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.