Abstract

In this article, design and finite element simulation of porous Ti-6Al-4V alloy structures was presented. Typically, titanium and titanium alloy implants can be manufactured with required pore size and porosity volume by using powder bed fusion techniques due to advancement in additive manufacturing technologies. However, the mismatch of elastic modulus between human cortical bone and the dense Ti-6Al-4V alloy implant resulted in stress shielding which accelerate the implant failure. The porous implant structures help in reduce the mismatch of elastic modulus between the cortical bone and implant structure and also improve the bone ingrowth. Hence, the present work focuses on design of Ti-6Al-4V alloy porous structures with various porosities ranging from 10% to 70% and simulated to determine the elastic modulus suitable for human cortical bone. The sample with 45% porosity is found to be best suited for replacement of cortical bone with elastic modulus of 74Gpa, preventing stress shielding effect and enhanced chances of bone ingrowth.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.